Over 15 years experience of providing support of MEAN WELL Power Supplies
Overview
GSM25U07-P1J is the 7.5V output model of this series: rated current 2.93A, typical efficiency 83%.
GSM25U is a highly reliable, 25W wall-mounted style single-output green medical adaptor series. This product is equipped with a 2-pin (no FG) standard USA AC power plug, adopting the input range from 80VAC to 264VAC. The entire series supplies different output voltages between 5VDC and 48VDC that can satisfy the demands for various kinds of miniaturize medical devices. The circuitry design meets the international medical standards (2×MOPP), having an ultra low leakage current (<50 μA), fitting the medical devices in direct electrical contact with the patients. With the efficiency up to 88% and the extreme low no-load power consumption below 0.1W, GSM25U is compliant with USA EISA 2007/DoE and Canada NRCan; the superme feature allows the adaptor to save the energy when it is either under the operating mode or the standby mode. The entire series utilizes the 94V-0 flame retardant plastic case, providing the double insulation that effectively prevents electrical shock. GSM25U is approved with the international medical safety certificates.
Features
- 2 pole USA plug, Class Ⅱ power unit
- Medical safety approved (2×MOPP) according to ANSI/AAMI ES60601-1/-1-11
- Extremely low leakage current
- No load power consumption<0.1W
- Energy efficiency Level Ⅵ ( except 5~9V for Level Ⅴ)
- -25~+60℃ wide range working temperature
- Protections: Short circuit / Overload / Over voltage
- LED indicator for power on
- Various DC plug quick adapter accessory available (Plug kit sold sperately, please refer to : https://www.meanwell.com/upload/pdf/DC_plug.pdf)
- 3 years warranty
Specifications
OUTPUT
- SAFETY MODEL NO.
- GSM25U07
- DC VOLTAGENote.2
- 7.5V
- RATED CURRENT
- 2.93A
- CURRENT RANGE
- 0 ~ 2.93A
- RATED POWER (max.)
- 22W
- RIPPLE & NOISE (max.)Note.3
- 80mVp-p
- VOLTAGE TOLERANCENote.4
- ±5.0%
- LINE REGULATIONNote.5
- ±1.0%
- LOAD REGULATION
- ±5.0%
- SETUP, RISE TIMENote.6
- 500ms, 30ms / 230VAC 1000ms, 30ms / 115VAC at full load
- HOLD UP TIME (Typ.)
- 16ms / 230VAC 16ms / 115VAC at full load
INPUT
- VOLTAGE RANGENote.7
- 80 ~ 264VAC 113 ~ 370VDC
- FREQUENCY RANGE
- 47 ~ 63Hz
- EFFICIENCY (Typ.)
- 83%
- AC CURRENT (Typ.)
- 0.7A / 115VAC 0.35A / 230VAC
- INRUSH CURRENT (Typ.)
- 55A / 230VAC 30A / 115VAC
- LEAKAGE CURRENT(max.)
- Touch current < 50μA/264VAC
PROTECTION
- OVERLOAD
- 105 ~ 170% rated output power Protection type : Hiccup mode, recovers automatically after fault condition is removed
- OVER VOLTAGE
- 7.88 ~ 10.5V Protection type : Shut down o/p voltage, re-power on to recover
ENVIRONMENT
- WORKING TEMP.
- -25 ~ +60℃ (Refer to "Derating Curve")
- WORKING HUMIDITY
- 20% ~ 90% RH non-condensing
- STORAGE TEMP., HUMIDITY
- -40 ~ +85℃, 10 ~ 95% RH non-condensing
- TEMP. COEFFICIENT
- ±0.03% / ℃ (0~40℃)
- VIBRATION
- 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
SAFETY & EMC (Note. 8)
- SAFETY STANDARDS
- IEC 60601-1:2005+A1+A2; IEC 60601-1-11:2015+A1, ANSI/AAMI ES60601-1:2005+A2; ANSI/AAMI HA60601-1-11+A1 CAN/CSA C22.2 No. 60601-1:2014+A2; CSA C22.2 NO. 60601-1-11:2015+A1, EAC TP TC 004 approved
- ISOLATION LEVEL
- Primary-Secondary: 2xMOPP
- WITHSTAND VOLTAGE
- I/P-O/P:4KVAC
- ISOLATION RESISTANCE
- I/P-O/P:100M Ohms / 500VDC / 25℃/ 70% RH
- EMC EMISSION
-
Parameter Standard Test Level Conducted emission FCC PART 15 / CISPR22 Class B Radiated emission FCC PART 15 / CISPR22 Class B
OTHERS
- MTBF
- 3887.1K hrs min. Telcordia SR-332 (Bellcore) ; 791.9K hrs min. MIL-HDBK-217F (25℃)
- DIMENSION
- 79*54*33mm (L*W*H)
- PACKING
- 215g ; 60pcs / 13.9Kg / CARTON
CONNECTOR
- PLUG
- See page 4~5 ; Other type available by customer requested
- CABLE
- See page 4~5 ; Other type available by customer requested
Test conditions and notes8
- 1. All parameters are specified at 115VAC input, rated load, 25℃ 70% RH ambient.
- 2. DC voltage: The output voltage set at point measure by plug terminal & 50% load.
- 3. Ripple & noise are measured at 20MHz by using a 12" twisted pair terminated with a 0.1μF & 47μF capacitor.
- 4. Tolerance: includes set up tolerance, line regulation, load regulation.
- 5. Line regulation is measured from low line to high line at rated load.
- 6. Length of set up time is measured at first cold start. Turning ON/OFF the power supply may lead to increase of the set up time.
- 7. Derating may be needed under low input voltage. Please check the derating curve for more details.
- 8. The power supply is considered as an independent unit, but the final equipment still need to re-confirm that the whole system complies with the EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.” (as available on https://www.meanwell.com//Upload/PDF/EMI_statement_en.pdf )
Technical drawings
Typical applications
- Blood glucose meter
- Blood pressure meter
- Nebulizer
- Inhaler
- Portable medical device
Downloads
Datasheet1
Certificates5
Test reports1
FAQ
Can I use the GSM25U07-P1J for 7.5V embedded or industrial control electronics?
Yes, if the equipment is designed for constant-voltage 7.5V input and stays within 22W total output. This model can provide up to 2.93A, which fits many compact embedded and control applications at 7.5V.
Is this 7.5V supply suitable for portable test equipment or monitoring devices?
It can be suitable for test or monitoring equipment that requires a stable 7.5V DC source and no more than 2.93A. The wide 80-264VAC input is also useful when the equipment may be used across different AC mains environments.
Can this model power a small automation interface that runs on 7.5V?
Yes, for automation interfaces or embedded devices built around a fixed 7.5V input and a load up to 22W. Because it is a constant-voltage supply, it is meant for electronics that specify a steady 7.5V source.
What is the practical load limit for this 7.5V power supply?
The practical limit is 22W total output, which corresponds to a maximum output current of 2.93A at 7.5V. Loads above 2.93A would exceed the rated capacity of this model.
Is the GSM25U07-P1J appropriate for outdoor or exposed environments?
Its operating temperature range is -25°C to +60°C, so the temperature spec covers many indoor equipment and control-panel conditions. However, no IP rating is listed in the provided specs, so use in exposed outdoor, wet, or dusty environments is not established here.
Payment & Security
Payment methods
Your payment information is processed securely. We do not store credit card details nor have access to your credit card information.
Other GSM25 Models
3 other models| Model | Power | Voltage | Current | IP Rating | Price | |
|---|---|---|---|---|---|---|
| Mean Well GSM25U24-P1J Power Supply 25W 24V 1.04A | 25W | 24V | 1.04A | IP20 | $18.95 | View |
| Mean Well GSM25U15-P1J Power Supply 25W 15V 1.66A | 25W | 15V | 1.66A | IP20 | $16.69 | View |
| Mean Well GSM25B12-P1J Power Supply 25W 12V 2.08A | 25W | 12V | 2.08A | IP20 | $16.69 | View |
Please log in to write a review.
Write a review
Reviews are checked before publishing. By submitting you confirm this is your genuine experience with this product.
Submitting your review…
It's published and now helps other buyers.
We're checking it and it'll appear shortly.
Please try again.