Mean Well RPS-65-15 Green Power Supply 65.1W 15V 4.34A - Medical Power Supply

3" x 2" compact sizemedical safety approvedno load power consumption < 0.1Wextremely low leakage current
SKU RPS-65-15
$17.67 USD
Shipping calculated at checkout.
|

Your order will be processed within 24 hours


Overview

RPS-65-15 is the 15V output model of this series: rated current 4.34A, rated power 65.1W, typical efficiency 89%.

RPS-65 is a 65W highly reliable green PCB type medical power supply with a high power density on the 3” by 2” footprint. It accepts 80~264VAC input and offers various output voltages between 3.3V and 48V. The working efficiency is up to 91% and the extremely low no load power consumption is down below 0.1W. RPS-65 is able to be used for ClassⅡ (no FG) system design. The extremely low leakage current is less than 100μA. In addition, it conforms to international medical regulations (2*MOPP) and EMC BS EN/EN55011, perfectly fitting all kinds of BF rated “patient contact” medical system equipment.

Features

  • 3"×2" compact size
  • Medical safety approved (2 x MOPP) according to ANSI/AAMI ES60601-1 and IEC/BS EN/EN60601-1
  • Suitable for BF application with appropriate system consideration
  • Cooling by free air convection
  • EMI class B for class Ⅱ configuration
  • No load power consumption<0.1W
  • Extremely low leakage current
  • Protections: Short circuit / Overload / Over voltage
  • Operating altitude up to 4000 meters
  • 3 years warranty

Specifications

RPS-65-15Specifications from the official datasheet (2026-10-09 edition)

OUTPUT

DC VOLTAGE
15V
RATED CURRENT
4.34A
CURRENT RANGE
0 ~ 4.77A
RATED POWER
65.1W
PEAK LOAD(10sec.)
71.6W
RIPPLE & NOISE (max.)Note.2
120mVp-p
VOLTAGE ADJ.RANGE
13.5~16.5V
VOLTAGE TOLERANCENote.3
±1.0%
LINE REGULATION
±0.5%
LOAD REGULATION
±1.0%
SETUP, RISE TIME
500ms, 30ms / 230VAC 500ms, 30ms / 115VAC at full load
HOLD UP TIME (Typ.)
30ms / 230VAC 12ms / 115VAC at full load

INPUT

VOLTAGE RANGENote.4
80 ~ 264VAC
FREQUENCY RANGE
47 ~ 63Hz
EFFICIENCY (Typ.)
89%
AC CURRENT (Typ.)
1.5A / 115VAC 1A / 230VAC
INRUSH CURRENT (Typ.)
COLD STAR 30A/115VAC 50A/230VAC
LEAKAGE CURRENT(max.)Note.5
Touch current< 100μA/264VAC

PROTECTION

OVERLOAD
115 ~ 150% rated output power Protection type : Hiccup mode, recovers automatically after fault condition is removed
OVER VOLTAGE
17.2~20.3V Protection type : Shut down o/p voltage, re-power on to recover

ENVIRONMENT

WORKING TEMP.
-30 ~ +70℃ (Refer to "Derating Curve")
WORKING HUMIDITY
20% ~ 90% RH non-condensing
STORAGE TEMP., HUMIDITY
-40 ~ +85℃, 10 ~ 95% RH non-condensing
TEMP. COEFFICIENT
±0.03% / ℃ (0 ~ 50℃)
VIBRATION
10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
OPERATING ALTITUDENote.6
4000 meters

SAFETY & EMC

SAFETY STANDARDS
IEC 60601-1:2005+A1+A2, TUV BS EN/ EN 60601-1:2006+A1+A12+A2, ANSI/AAMI ES60601-1:2005+A2 CAN/CSA C22.2 No. 60601-1:2014+A2, EAC TP TC 004 approved; Design refer to BS EN/EN60335-1(by request)
ISOLATION LEVEL
Primary-Secondary: 2xMOPP
WITHSTAND VOLTAGE
I/P-O/P: 4KVAC
ISOLATION RESISTANCE
I/P-O/P:100M Ohms / 500VDC / 25℃/ 70% RH
EMC EMISSION
ParameterStandardTest Level
Conducted emissionBS EN/EN55011 (CISPR11)Class B
Radiated emissionBS EN/EN55011 (CISPR11)Class B
Harmonic currentBS EN/EN61000-3-2Class A
Voltage flickerBS EN/EN61000-3-3
EMC IMMUNITYNote. 7

BS EN/EN55035,BS EN/EN60601-1-2

ParameterStandardTest Level
ESDBS EN/EN61000-4-2Level 4, 15KV air; Level 4, 8KV contact
RF field susceptibilityBS EN/EN61000-4-3Level 3, 10V/m( 80MHz~2.7GHz ) Table 9, 9~28V/m( 385MHz~5.78GHz )
EFT burstsBS EN/EN61000-4-4Level 3, 2KV
Surge susceptibilityBS EN/EN61000-4-5Level 4, 2KV/Line-Line
Conducted susceptibilityBS EN/EN61000-4-6Level 3, 10V
Magnetic field immunityBS EN/EN61000-4-8Level 4, 30A/m
Voltage dip, interruptionBS EN/EN61000-4-11100% dip 1 periods, 30% dip 25 periods, 100% interruptions 250 periods

OTHERS

MTBF
3334.3K hrs min. Telcordia SR-332 (Bellcore) ; 959.1K hrs min. MIL-HDBK-217F (25℃)
DIMENSION (L*W*H)
76.2*50.8*24mm or 3" * 2" *0.945" inch
PACKING
0.11Kg; 120pcs/14.2Kg/0.94CUFT
Test conditions and notes7
  1. 1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25℃ of ambient temperature.
  2. 2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1μF & 47μF parallel capacitor.
  3. 3. Tolerance : includes set up tolerance, line regulation and load regulation.
  4. 4. Derating may be needed under low input voltages. Please check the derating curve for more details.
  5. 5. Touch current was measured from primary input to DC output.
  6. 6. The ambient temperature derating of 3.5℃/1000m with fanless models and of 5℃/1000m with fan models for operating altitude higher than 2000m(6500ft).
  7. 7. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.” (as available on https://www.meanwell.com//Upload/PDF/EMI_statement_en.pdf )

Technical drawings

RPS-65-15 Block Diagram
Block Diagram
RPS-65-15 Mechanical Specification
Mechanical Specification

Typical applications

  • Oral irrigator
  • Hemodialysis machine
  • Medical computer monitors
  • Sleep apnea devices

Downloads

FAQ

Can I use the RPS-65-15 for 15V industrial instrumentation or control electronics?

Yes, if the equipment needs a constant 15V output and draws no more than 4.34A, with 65.1W total power available. It is intended for fixed-voltage DC applications rather than constant-current loads.

Is the RPS-65-15 a good fit for 15V embedded or OEM systems?

It can be a good fit for embedded or OEM assemblies that require a regulated 15V rail within 4.34A. The compact 76.2 x 50.8 x 24.0 mm footprint and 0.11 kg weight make it easier to place in space-constrained equipment.

Can this model be used in medical or diagnostic equipment that operates at 15V?

This supply is medical safety approved and specified with extremely low leakage current, which is relevant for medical and diagnostic equipment selection. It is appropriate where the system needs 15V constant voltage and stays within 65.1W and 4.34A.

Would the RPS-65-15 suit low-standby-power devices or instruments?

The listed no-load power consumption is below 0.1W, so it suits designs where idle consumption is a consideration. During operation, it can supply 15V at up to 4.34A.

Can the RPS-65-15 be used in outdoor control cabinets?

It is rated for operation from -30°C to +70°C, so the temperature range is specified. However, no IP rating is provided, so outdoor suitability depends on the cabinet or enclosure rather than the power supply alone.

Payment & Security

Payment methods

  • Amazon
  • American Express
  • Apple Pay
  • Bancontact
  • Bizum
  • Diners Club
  • Discover
  • Google Pay
  • iDEAL Wero
  • Mastercard
  • PayPal
  • Satispay
  • Shop Pay
  • USDC
  • Venmo
  • Visa

Your payment information is processed securely. We do not store credit card details nor have access to your credit card information.

Mean Well RPS-65-15 Green Power Supply W 15V 4.34A - Medical Power Supply - RPS-65-15 - powersupplymall.com
Mean Well

Mean Well RPS-65-15 Green Power Supply 65.1W 15V 4.34A - Medical Power Supply

$17.67 USD

RPS-65-15 is the 15V output model of this series: rated current 4.34A, rated power 65.1W, typical efficiency 89%.

RPS-65 is a 65W highly reliable green PCB type medical power supply with a high power density on the 3” by 2” footprint. It accepts 80~264VAC input and offers various output voltages between 3.3V and 48V. The working efficiency is up to 91% and the extremely low no load power consumption is down below 0.1W. RPS-65 is able to be used for ClassⅡ (no FG) system design. The extremely low leakage current is less than 100μA. In addition, it conforms to international medical regulations (2*MOPP) and EMC BS EN/EN55011, perfectly fitting all kinds of BF rated “patient contact” medical system equipment.

Features

  • 3"×2" compact size
  • Medical safety approved (2 x MOPP) according to ANSI/AAMI ES60601-1 and IEC/BS EN/EN60601-1
  • Suitable for BF application with appropriate system consideration
  • Cooling by free air convection
  • EMI class B for class Ⅱ configuration
  • No load power consumption<0.1W
  • Extremely low leakage current
  • Protections: Short circuit / Overload / Over voltage
  • Operating altitude up to 4000 meters
  • 3 years warranty

Typical applications

  • Oral irrigator
  • Hemodialysis machine
  • Medical computer monitors
  • Sleep apnea devices
View product